PET Cyclotron

PET Cyclotron

CYPRIS HM-12

 

   

                       HM-12MeV Medical Cyclotron

Sumitomo HM-12 Cyclotron using sate-of-the-art technology realizes high yield performance and compact body size at the same time.HM-12 is dual-beam

cyclotron to generate 12 MeV proton beam and 6 MeV deuteron beam as well.

Production of oxegen15 is available by deuteron irradiation to 14N-N2 gas without using enriched 15N-N2 gas, which leads to a substantial reduction of operation costs. The compact target system is designed for attaching 4 targets per each beam port (total 8 targets).

CHARACTERISTICS OF HM-12

Particle

Proton

Deuteron

Energy

12 MeV

6MeV

No. of Targets

8

Dimension(W)*(D)*(H)

2.4m*1.6m*1.8m

 

(S)3.95m*2.45m*2.25m

 

 

CYPRIS HM-20

                                                HM-20MeV Medical Cyclotron

Sumitomo HM-20V cyclotron is a fixed-energy isochronous cyclotron, which accelerates negatively charged hydrogen ions (H-) to 20 MeV and (D-) to 10MeV. The total extracted beam current (target plus collimator currents) is 150 µA for proton beam and 50 µA for deuteron beam respectively.This cyclotron has a very high degree of automation built into the system. It can be equipped with up to 8 (eight) targets and various chemistry systems for labeling of the most common radioisotopes used in Positron Emission Tomography (PET) systems.

 

 

CHARACTERISTICS OF HM-20

Particle

Proton

Deuteron

Energy

20 MeV

10MeV

No. of Targets

8

Dimension(W)*(D)*(H)

2.6m*1.7m*2.0m

 

(S)4.5m*3.0m*2.7m

FDG Auto Dispensing System

FDG Auto Dispensing System

Specification-

  1. Dispense Mode: Request dispensing/Full product washing out
  2. Maximum input activity : 55.5 GBq
  3. Maximum dispense Activity : 18.5 GBq (for FDG Auto Injection System)
  4. Minimum dispense Activity : 185 MBq (using retract function)
  5. Dispense RI accuracy : +- 10% (with maximum input)
  6. Dispense Volume accuracy : +- 10%(with maximum input)
  7. Dispense Interval : 3 min

18O Water Purifier

18O Water Purifier

Feature-

This system consists of 2 units, i.e.

  1. Photox-unit, Evap-Cond unit.
  • Normal capacity is 16gr/working day
  • Photox Unit: Dimensions (W x D x H): 400x300x350mm
  • Components: Quartz cells(10 pieces) with caps, a cell rack a blower and an ozone eliminator

        2.   Evap-Cond Unit:

  • Dimensions ( W x D x H) : 300x300x500mmComponents:
  • Evaporation glass ware & pot, a dewer flask, polystyrene tubes and a diaphragm vacuum

Auto Injector AI300

Auto Injector AI300

This is equipment for administering 18F-FDG radioactive medications to test subjects.

AUTO QC FOR FDG

Auto QC for FDG

Feature

  1. Automatic Quality check  system has developed to avoid radiation exposure
  2. Quality record is stored in electronic form and unrewritable to match cGMP requirement.